B3 radiochromic dosimeters

Simple hand-held dosimeter with unique barcode. Small, versatile and robust. Ideal for most industrial gamma, e-beam and x-ray radiation applications, including routine dosimetry, testing, qualification and dose mapping.

Proven reliability and used in commercial industrial applications worldwide since 1991.

B3 Benefits
  • Versatile, cost-effective dosimetry
  • Easy for operators to use
  • Consistent, traceable results
  • Robust and ideal for manufacturing environments
B3 Overview
  • Dose range of 1.0 kGy to 150 kGy in Gamma, Electron Beam, and X-ray.
  • Absorbed dose rates > 0.01 Gy/s.
  • Photon energy range approximately 0.1 to 50 MeV.
  • Electron energy range is approximately 70 KeV to 50 MeV.
Applications for B3
  • High energy 10MeV electron beam
  • Low energy 100-200 keV electron beam
  • Gamma
  • X-ray
  • Routine dosimetry, process control and IQ/OQ/PQ
check_circle

General FAQ's

Answers to general questions about B3 film and B3 dosimeters.

Click here.
star

Spec and Usage FAQ's

More detailed answers to questions about how B3 dosimeters work and how to use the products.

Click here.
verified

Compare B3 to other dosimeters

How B3 dosimeters compares with alanine, PMMA (Harwell) and FWT dosimeters.

Click here.

Frequently Asked Questions

About B3 film and dosimeter products

B3 is the trade name of a thin radiochromic film material used for measuring the radiation dose used in industrial processes. GEX Corporation uses the B3 radiochromic film in the manufacturing of its proprietary B3 dosimeters.

The B3 film is pararosaniline cyanide dissolved in polyvinyl butyral (PVB), and is manufactured into a highly-uniform, thin film with a nominal 18μm thickness.

Upon irradiation B3 film undergoes a predictable and quantifiable chemical color change directly related to the amount of dose of ionizing radiation energy it receives.

B3 dosimeters provide consistent, accurate and precise dose measurements. B3 dosimeters are stable after irradiation by applying a heat treatment.

As with all routine dosimeters, calibration should be carried out with the user’s specific instruments, environment and practices with dosimeters irradiated under conditions of actual use.

For routine use dosimetry systems, wide response range allows it to be calibrated from below 1.0 kGy to more than 100 kGy.

When used as a reference dosimetry system, the useful range of B3 is limited with a recommended calibration range of approximately 5.0 kGy to 30 kGy.

B3 film can be used to measure ionizing radiation doses from gamma, X-ray and electron beam sources including low energy electron sources of 80 keV or lower.

B3 dosimeters are used for measuring the radiation dose used in industrial processes, such as medical device sterilization, surface sterilization, crosslinking and curing of inks and coatings, and food irradiation. B3 dosimeters are not used for personnel monitoring, radiation safety, or nuclear power industries.

B3 dosimeters can be used to measure ionizing radiation doses from gamma, X-ray and electron beam sources, including low energy electron sources, with a dose range from <1.0 kGy to > 150 kGy.

B3 film is ideally measured in a spectrophotometer, to determine the optical absorbance after irradiation. There is no need to measure the individual thickness of B3 dosimeters because of the consistent uniformity of the B3 film.

The spectrophotometer measurement method is non-destructive so dosimeters can be re-measured. B3 dosimeters are stable for a long time after irradiation if the dosimeters are stored under normal laboratory conditions.

GEX recommends the Thermo Evolution and Genesys 30 spectrophotometers, included as part of the complete DoseControl Dosimetry System.

B3 film can also be measured using specialized software called RisøScan. RisøScan B3 dosimeter measurement software allows users to scan and analyze B3 dosimeter film using high-resolution bitmap scans from a flatbed scanner.

B3 radiochromic dosimetry has been heavily characterized over the commonly used application ranges with published data references dating back to the 1970s.

Each batch of B3 dosimeters is characterized for the repeatability of specific net absorbance by analyzing the data from the sets of dosimeters irradiated during the calibration at each dose value.

Appropriate documented information regarding the effects of temperature, humidity, UV light exposure, and post-irradiation stability of B3 dosimeters are described in the following publication:

Abdel-Fattah A. A. and Miller A. (1996); Temperature, Humidity, and Time. Combined Effects on Radiochromic Film Dosimeters; Radiation Physics and Chem. Vol. 47, No. 4 pp 611-621, Elsevier Science Ltd. Pergamon Press, Great Britain.

Contact GEX for a concise set of recommended practices or rules to use with B3 radiochromic film dosimetry in order to obtain optimum performance results and discuss your specific dosimetry requirements.

The B3 dosimeter is the preferred product choice for most applications and dose ranges from <1.0kGy to >150kGy used in commercial routine irradiation processes. Our alanine dosimetry option is an excellent product choice for low dose applications below 5.0kGy.

When choosing a dosimetry system for you application, an appropriate dosimeter should:

  • Be able to measure the full range of dose to be applied in routine irradiation product processing, equipment characterization, and process testing.
  • Have the physical dimensions and characteristics necessary for successful measurement results.
  • Be versatile and able to be securely fixed against the product surface(s) or wherever dose is to be measured.

GEX technical experts can guide you through the selection process, and can define a solution that works for your specific application requirements.

B3 is well characterized and established for use in virtually all current radiation processing applications, provided it is calibrated under the approximate conditions of actual use. Use outside of the noted dose range is possible but may have limitations.

Benefits and Advantages of B3 dosimeter products
  • Superior measurement precision and accuracy. B3 film thickness uniformity results in significantly lower dose measurement variability.
  • Stability. B3 dosimeters are completely stable before and after irradiation.
  • Usability.  B3 dosimeters are thin enough for use in all radiation processing application, highly flexible, and can be placed directly on products and readily conform to irregular surface geometries.  

B3 dosimeters are quick and simple to use, easy to integrate into a manufacturing process, and easy to integrate into a software database solution.

  • B3 dosimeters are barcoded to provide faster readability. All doses read by the spectrophotometer for each dosimeter can be directly imported into a database.
  • B3 dosimeters are significantly faster to read when compared to other systems.
  • B3 dosimeters are thin, flexible and are featured in an array of dosimeter products designed for specific measurement activities.

B3 dosimeters provide unsurpassed quality assurance

  • The B3 dosimeter exceeds standards of accuracy and precision that is set to meet the stringent dosimetry demands of the sterilization and food irradiation applications.
  • Due to the B3’s material uniformity, there is significantly lower dose measurement uncertainty.
  • B3 dosimeters are completely stable before irradiation, and can be used right off the shelf. 
  • Because the B3 dosimeter is thin, it can be used in a wide variety of electron beam processes.
  • All B3 dosimeters are traceable to the original source and placement in the manufacturing process, because each dosimeter is barcoded and numbered.

B3 dosimeters are a cost-effective solution

  • Because B3 dosimeters have such a high thickness uniformity and are stable before irradiation, there is no need to have to measure individual thickness or monitor pre-irradiation background absorbance values, saving valuable time.
  • Because B3 dosimeters can be measured quickly, this saves related labor costs.

We provide a detailed comparision of B3 dosimeters to alanine dosimeters on this page below: B3 Film compared to Alanine - click here.

We provide a detailed comparision of B3 dosimeters to FWT-60 dosimeters on this page below: B3 Film compared to FWT-60 - click here.

Yes. Complete B3 Dosimetry System implementation and validation guides and documentation are standard with the purchase of B3 products. GEX offers comprehensive customer support for system implementation and validation.

GEX also offers an on-site service IQ/OQ for dosimetry systems, for an additional charge. Dosimeter batch calibration services are also available. For more information about the On-site IQ/OQ Service or Dosimeter Batch Calibration services, please contact sales here.

History and Development of B3 dosimeter products

Arne Miller, Director of the High Dose Reference Laboratory at Risø National Laboratory, now the Danish Technical University (DTU), was responsible for the development of B3 radiochromic film in 1976. In an effort to create a next-generation radiochromic dosimeter with improved stability and other advancements, Mr. Miller collaborated with Dr. William McLaughlin, of NIST, who originally identified and investigated the potential of radiochromic film as a dosimeter material.

B3 film was developed as a next-generation radiochromic film dosimeter, with a wide dose range, measured at a single wavelength, featuring stability and measurement accuracy advantages over existing radiochromic film dosimeters.



B3 Specifications and Usage

Quick answers to common questions about B3 specifications and usage

Type: Spectrophotometer

Dosimeter Holder: The use of GEX specially designed holders is recommended to orient the film perpendicular to the light beam in the spectrophotometer. Holders must prevent the interference of the paper frames with the spectrophotometer light beam when using B3 Dosimeters.

Spectral Bandwidth (SBW): 5nm or greater. Best performance is with 8nm or better. This is required to reduce variability from optical interference fringing.

Wavelength Setting: The absorbance wavelength peak of B3 film is 552 nm and the dosimeters should always be measured at 552nm.

Photometric Range:0.00 to 2.00 or higher

Measurement of optical absorbance is affected by the optical path-length (thickness) of the sample. Therefore, the thickness variance between dosimeters contributes to the variability of the measurement of absorbance between samples.

GEX B3 dosimeters are manufactured in large, highly uniform batches that allow users to apply an average dosimeter thickness correction in the measurement process. A specified average thickness and variability is measured, assigned and printed on each box of GEX B3 dosimeters and in a Certificate of Compliance to the user with each order. It is recommended to use the certified thickness rather than attempting to measure thickness.

Users can measure thickness although it is not necessary or beneficial to do so. Caution in the selection of thickness measurement device is advised because the film is very thin. Mechanical gauges are not precise enough to resolve the thickness better than ±0.3 microns. Therefore, GEX suggests the use of an optical thickness gauge for those users who want to measure thickness.

General Information: The response function of radiochromic film dosimeters is affected by the temperature and the water content in the film material at the time of irradiation. These are the two most important influence quantities. B3 dosimeter products are typically supplied in sealed pouches to maintain the specific controlled environment in which they were packaged and sealed. The films should be used in factory sealed packaging whenever possible.

However, the films must be used without packaging by many users and the user is cautioned to apply appropriate techniques to verify the validity of measurements under such conditions against a calibrated reference.

The best method for controlling the effect of influence quantities is to calibrate the dosimeters under the conditions of use. This includes using the packaging, environmental conditions, post-irradiation treatment as well as the procedures and methods of routine operation during the calibration of the dosimetry system. (See section on Calibration for more information).

Pre-Irradiation Storage: Store unopened packages under ambient conditions (15°C to 30°C). Exposure to temperatures outside this range is acceptable, but the film should not be exposed to temperatures above 45°C. Exposure above 45°C will permanently change the response function of the film.

The packaging of B3 dosimeters is designed to provide an environmentally stable condition for the dosimeters when stored under the above conditions. The shelf life for B3 dosimeter packages is 5 years, and 3 years for B3DoseStix dosimeter packages, (refer to the Certificate of Compliance that accompanies each shipment). The film itself does not have a known shelf life if calibrated at regular intervals.

Dose rate effects: There are no known dose rate effects.

Irradiation Temperature: Caution: Use of B3 Film products in irradiation temperatures above 60°C are at the user’s discretion per ISO/ASTM 51275. Irradiation above 60°C generally causes a reduction of the response function and more variable measurements.

Relative Humidity: The water content in the film affects the response. Use factory sealed packaging or verify the validity of measurements made without packaging using a reference. Humidity extremes <10% or >90% may cause more variation in measurement.

UV Exposure: The B3 radiochromic film is sensitive to UV radiation. Therefore, control exposure of unprotected film dosimeters to daylight and fluorescent light sources. UV filters on lighting and windows in the measurement area are recommended.

Post Irradiation Stability: The coloration of B3 radiochromic film may continue to develop for some period of time after irradiation. This is known as the influence of time on the response. To eliminate dosimeter response variance, post-irradiation heat treatment of B3 film dosimeters is recommended to complete the color development cycle and render B3 dosimeters stable. The user should validate the heat treatment method used and consult GEX literature on the subject for guidance.

Calibrate under conditions approximating actual usage. For guidance, see: ISO/ASTM 51261 and NPL Report CIRM 29.

General: The dosimeter batch calibration should be accomplished under normal process (including environmental) conditions. Dosimeters should be measured at a time interval(s) consistent with routine production dosimetry.

Dose Range: 1.0 kGy - 150 kGy or as determined by user’s application and uncertainty requirements. Using multiple replicate measurements or calibrating multiple dose ranges can result in significant reductions in overall uncertainty.

Calibration Geometry: Use an appropriate calibration phantom (as described in ISO/ASTM 51261 and NPL CIRM Report 29) to co-locate reference-transfer standard dosimeters in controlled proximity to the routine dosimeters being calibrated.

Learn more about GEX dosimeter batch calibration service.

Post-irradiation heat treatment of B3 film is necessary to catalyze chemical reaction and finalize the color change. Heat treatment is required for all users except those users exclusively using high doses (> 40 kGy) and high temperatures (> 55.0°C). For complete information about heat treating B3 dosimeters, reference GEX TIR #100-201.

  • GEX forced air incubator - GEX product #P4800 - The correct minimum heat treatment time in an oven-type incubator is 15 minutes at 60C for the B3 dosimeter in the pouch.
  • For GEX product #P4900 Microincubator, 5 minutes at 58.5C, for B3 dosimeter in the pouch.

Guidelines for time interval between dosimeter irradiation and dosimeter heat treatment

  • It is critical that the time interval used during the calibration of each B3 film dosimeter batch be as similar as possible to the process used for measuring routinely irradiated samples and the user should quantity any major differences.
  • o The time interval should be estimated from the time the dosimeter leaves the irradiation zone and not just the time since it is retrieved from the process. For optimal results heat treat dosimeters at least 1 hour after irradiation for Ebeam processes to allow the dosimeter to naturally respond to the massive amount of energy that was imparted by the beam. Some users may not be able to wait 1 hour and should test the effects of a shorter interval. This guidance is a simple answer to a complex issue - users should test within the user's process, reference GEX TIR #100-201.

Caution: There is a risk of damage to B3 film if any material contacts the film, including another B3 dosimeter, when the film surface temperature exceeds the (Tg) of PVB (1 minute of film surface temperature in excess of 58.0°C ). The risks increase with every degree above it that the user employs.

Do not touch the film. Use powder free gloves or handling tools. Static conditions may cause particulate to be drawn to the dosimeter surface which can cause readout error. Remove visible particulate with a soft brush or piece of soft anti-static cloth. Gently tapping the dosimeter can also dislodge particulate.

A batch ID and unique dosimeter identification number is found on each B3 dosimeter. This provides users with absolute dosimeter verification information. Dosimeter thickness information may be found on the product box.

B3 dosimeters contain barcodes that can be used to provide automated data entry of the unique dosimeter ID numbers directly to computer based data tracking and dosimetry software systems. The B3DoseStix and new B3 Dosimeter barcodes can be scanned through the packaging window.

B3 dosimeters have a perforated seam on the paper in between the barcoded paper flag and the dosimeter film area. You can tear the dosimeter into a small, square shape without damaging the film. This makes it easier to perform dose mapping activities in small spaces.



B3 dosimeters compared to alanine dosimeters

NOTE: The comparison below represents GEX's opinion and there may be other opinions that are not represented. GEX admits that some subjectivity is required to make an assessment, and the statements made are subjective based on GEX's analysis of the dosimeters.
Characteristic B3 Radiochromic Alanine Choice
Measurement Without question the B3 measurement process is faster and simpler, however, the user must remove the dosimeter from the package which then exposes the dosimeter to different RH. In a perfect world, measurement in the package would be preferred if it were possible. Still, all in all the B3 is simpler although this is subjective and biased because of our breadth of experience with B3. Someone may disagree with this more than other categories discussed herein. Measurement is significantly longer and more complicated, however, the dosimeter is measured in the package which is very convenient. The sample can be stored and remeasured with destroying the environment, which avoids adding additional variable for reread measurements.

done B3

Influence of quantities during irradiation such as temperature, RH, and dose magnitude on the response. The B3 film is affected by temperature and RH during irradiation in a complex, interdependent manner that varies with the magnitude of dose. For processes that are highly variable with respect to these factors, such as contract sterilization using gamma, the B3 will not perform as well as alanine, because it is impossible to capture the variety of these conditions in the calibration exercise. Alanine is affected by irradiation temperature at a rate between 0.11% and 0.15% of Hpp per °C. This is captured in the calibration and will result in variation in dose only when the actual conditions vary from the calibration conditions, but when it does the effect is comparably small compared to all radiochromic type dosimeters which can be affected at a rate 10 times that amount.

done Alanine

Spatial Resolution in Ebeam All dosimeters report dose to the dosimeter not to the product. The way we infer dose to the product in ebeam is by getting the dosimeter into position as a surrogate for the product surface. This cannot be done with alanine pellets. The majority of the surface area of an alanine pellet is not in contact with the product surface. Attempting to measure dose to product with alanine pellets will lead to under reporting actual dose due to the lack of scattering from the air around the pellet. This can lead to a distorted understanding of dose to product. Alanine pellet use in this manner is not recognized by industry experts as a valid method in electron beam dosimetry.

done B3

Dosimeter Homogeneity 18 micron thin film are not simple to produce, much less highly uniform film that isn’t produced by the tens of kilometers per week. Approximately 1200 m2 is produced every 1.5 years, not all of which is acceptable. This is sufficient to manufacture in excess of 2 million dosimeters but broken into 5 or 6 groups of thickness +/- 0.3 microns or less at k=2. There is no way to directly measure the impact of dosimeter homogeneity on performance. However, we know that optimal results occur if every dosimeter is identical to the other. Alanine is a very simple compound that requires good mixing of two key ingredients: the amino acid and the paraffin binder. The manufactured pellets are highly uniform and then are further sorted by weight. The largest variance in prepackaged alanines is probably the packaging itself which does affect the response measurement. In comparing the manufacturing complexity and results alanine is slightly more uniform, in general, due to its simplicity.

done Alanine

Instrument performance verification (instrument calibration) Spectrophotometer performance verification is well-understood and very good standards and methods exist to verify instrument performance. The user can define the intervals based on their risk-level but rest assured in the results without a lingering concern about performance. One of the major issues with alanine is that there is no reference standard to verify the stability of the Hpp signal over time. Is the instrument performing the same today as yesterday or last month? Rubies work well but they are not a CRM. Manufacturers have done various things using mfg. designed references and processes but the allowable variance from the manufacturer for these P.V. methods is quite high comparable to spectrophotometry. Plainly put the system is not as industrialized and standardized because of lack of a strong reference material to use for ensuring ongoing instrument performance.
For this reason and the sensitivity of ESR equipment in general, the system is less robust without significant monitoring efforts by the user.

done B3

Post irradiation stability If the user is able to follow a post irradiation heat treatment protocol with minor variance in the time between irradiation and heat treatment (compared to protocol during dosimeter batch calibration) then the B3 is more stable than any dosimeter. However, this comes at the cost of maintaining a heat treatment process.
Once heat treated the B3 is more stable than alanine (<1% drop in signal per year)
One of the major issues with alanine is that there is no reference standard to verify the stability of the Hpp signal over time. Is the instrument performing the same today as yesterday or last month? Rubies work well but they are not a CRM. Manufacturers have done various things using mfg. designed references and processes but the allowable variance from the manufacturer for these P.V. methods is quite high comparable to spectrophotometry. Plainly put the system is not as industrialized and standardized because of lack of a strong reference material to use for ensuring ongoing instrument performance.
For this reason and the sensitivity of ESR equipment in general, the system is less robust without significant monitoring efforts by the user.

done Alanine


Comparing B3 dosimeters to FWT-60 dosimeters

Many companies that use radiation sterilization with gamma or ebeam have chosen to convert their dosimetry from FWT-60 to GEX B3 film over the past 20 years. While both dosimeters share a lot of characteristics, there are some key reasons why some companies have made this decision. The information below outlines the major points of technical consideration.

We have learned from these user conversions many things about the B3 and our focus is on our product. We don’t test FWT product and rely on the input from the companies that have converted. Except for the world’s largest user of FWT-60 based in the USA, GEX has converted >95% of the remaining US market for film dosimetry in gamma and ebeam to B3. In addition, GEX B3 is used by companies, institutes, and laboratories in more than 55 countries around the world. GEX sells direct to most countries but has distribution networks in China, Japan, and France and prefers to work directly with its users.

Batch Size and Manufacturing Methods

GEX B3 dosimeters utilize Risø B3 film that are laminated to paper, serialized and in some cases barcoded, and packaged. The film is manufactured using solvent film casting process in large quantities of 1000-1500m2 that is produced over 3 days resulting in 3 large batches. The manufacturing is conducted on a machine that coats the film onto a moving web of backing material that is then dried in a highly controlled 30m drying tunnel. The resulting film has very few, if any, areas of any significant thickness variation. The film is then characterized for thickness and response uniformity, any areas of defect removed, and the batch is sub-grouped into thickness based on a normal distribution with the majority (90%) of the film being grouped into 3 thickness groups that are ±0.0003 mm or less.

By contrast, FWT film is historically produced and cast by hand and allowed to air dry. It is then cut to size and each film is measured individually. This process generally results in higher variance of the film thickness over any small area due to the fact of the air drying. In addition, the method of individually measuring and sorting the films which is necessary due to the means of production has in the past produced a certain percentage of outliers.

Characteristic B3 Radiochromic FWT-60
Type Radiochromic Film Radiochromic Film
Composition pararosaniline cyanide dye dissolved in polyvinyl butyral (PVB) hexa pararosaniline dye dissolved in nylon
Dimensions 10 mm2 (WINdose)
15mm x 50mm (DoseStix**)
10 mm2
Nominal Thickness 0.0180 mm 0.0500 mm
Analysis Wavelength 552 nm 605 nm (primary)
510 nm (secondary)
Dose Range++ 1 - 150 kGy 5 - 200 kGy (510 nm)
1 - 80 kGy (605 nm)
Reference Document ISO/ASTM-51275 ISO/ASTM-51275
Calibration Method In-situ using transfer-standard alanine for reference In-situ using transfer-standard alanine for reference
Meausrmenet Uncertainty (k=2) 4-6% 5-7%
Available Barcoded Pouched Dosimeters yes yes
Serialized, individually numbered dosimeters yes no
Barcoded dosimeters available? yes - DoseStix
available 2022 - new B3 Dosimeter
no
Pre Irradiation Signal Stability 100% stable Marginally unstable. User manages slow continual increase.
Post Irradiation Signal Management Heat Treatment* Time Management or Heat Treatment
Total Measurement time <5s <5s
Total Measurement time <5s <5s

* Heat Treatment to catalyze and complete the reaction renders films 100% stable for 30 days or longer.

** The dimensions of the film area are approximately 15mm x 13mm.

++ Both dosimeters may have the possibility to be used outside of these ranges but these numbers are generally accepted values.

Response Sensitivity

The FWT-60 measured at 605nm is far superior in accuracy than measured at 510nm. It has a high sensitivity which increases the accuracy of the system by reducing the impact of small changes in absorbance on the dose. This allows for more fluctuation of the signal (approximately ±0.004 A) for every 0.1 kGy. This is wonderful and provides for greater sensitivity at the low end of the dose range capability. However, the steep slope also is highly dependent on the consistency of the dose from the calibration laboratory during the fitting process. Small changes in that dose can realize much larger changes in the calibration function. Given that most labs are 2-3% uncertainty, this can cause shifting to occur due solely to the lab dose movement from one calibration to the next.

The B3 has a balanced sensitivity by design and one wavelength of measurement used over the full range of measurement. This means that the response is not as sensitive to very small changes in dose; it has a lower resolution than FWT and should only be used to calculate dose to one/tenth (0.1) kilogray maximum. However, it is not as sensitive to changes from the laboratory because the slope of the curve is about 1:1 in the middle of the medical device processing dose range.